  • JK-8CMultiplex temperature tester - copy - copy
JK-8CMultiplex temperature tester - copy - copy

JK-8CMultiplex temperature tester - copy - copy

JK-8C multi-channel temperature tester

The difference between the 3 multi-channel temperature tester

Model

300V
live test

Internal storage

U disk storage port

RS232
communication interface

USB
communication interface

software kit

JK-XU

2004 edition(latest)

JK-XA

2000 edition

JK-XC

2000 edition



JK-8C multi-channel temperature tester equipped with large screen LCD display, U disk interface, The utility model can conveniently measure and display the temperature data, and is an instrument which can be used for real-time monitoring and tracking at the same time. The high voltage semiconductor relay is used to scan the input signal to realize the high speed scanning, which eliminates the noise and the service life of the relay.

Technical characteristics:

Can set the upper and lower limits of the temperature value of each channel, beyond the sound alarm. Large screen liquid crystal display, can also display the multi-channel temperature value with U disk interface, insert the U disk can be massive preservation of temperature recording data. Especially for the oven, sterilization cabinet, lamps, household appliances, electric motors, electric appliances, temperature control devices, transformers, thermal protectors, and other manufacturers of the industry and the quality inspection department of multi-point temperature field detection. Energy saving lamp, electronic ballast inside the triode, choke coil, magnetic ring, capacitors and other components of the temperature at the same time live real-time monitoring, The whole process of temperature change is recorded by the computer in a curve way, which is convenient for analysis and improvement, and plays an important role in improving the reliability of the energy saving lamp and the electronic ballast.

model

channel

Live test

Internal storage

U disk storage port

RS232 communication interface (old)
USB communication interface 口(new)

JK-XC

8

Support 300V

yes

yes

yes

16

Support 300V

yes

yes

yes

24

Support 300V

yes

yes

yes

32

Support 300V

yes

yes

yes

64

Support 300V

yes

yes

yes


Main technical index:

1, temperature range: -100℃ ~ 1000℃;

2, measuring accuracy: 0 ~ 1000 DEG ℃: + (read value * 0.5%+1) ℃, -100 ~ 0℃: + (read value * 0.5%+2) ℃;

3, with high frequency interference function.

4, temperature signal input channel number: up to 8 groups, each group of 8 road;

5, sensor: nickel chromium nickel silicon (K) thermocouple (type T, type J can be specially made).

Permissible environmental conditions

1, power supply: AC 220V + 10%, 50Hz + + 2%;

2, the use of the environment: temperature: -2070 temperature, relative humidity: 20%90%;

3, the highest potential difference between thermocouples: 350V (true RMS);

volume

1, external dimensions: (L * w * h) 36cm * 26cm * 16cm

2, the whole machine weight: About 5kg


Multichannel temperature measuring instrument:
1: multi-channel temperature testing software and instrument with the use of the computer automatically receive test data, to achieve temperature records, storage and curve display, and print tables and curves.

2: JK-XC parameters can be set to choose any 1~64 path measurement, with different thermocouple to meet the different temperature measurement range; large screen LCD display, the temperature parameters and temperature curve display,(Different temperature All display and temperature alarm channel)Computer serial communication is directly connected to the computer to collect data, the software is specifically for the user to use the special requirements of multi-channel temperature, The utility model can be used for simultaneously measuring a sample or a plurality of different samples, or recording the names of a plurality of test sites, or measuring a sample or a plurality of different samples in different sizes, Can set the color temperature curve, there are professional design simplified Chinese and English version, suitable for a variety of different test needs

3: cold end ambient temperature: measuring range 0 ~ 60 C

4: the machine also has a date and time set with fixed point, patrol function; inspection time can be set: 1 to 255 seconds, the measurement time interval can be set (not less than 1 second);

5: temperature unit Celsius (C) Fahrenheit (F) can be automatically converted; an instrument can choose T J K thermocouple

6: test channel can be set to meet the measurement requirements and speed;

7: status bar displays the setting parameters in real time; the channel nonuniformity: 0.5℃

8: thermocouple probe can contact with no more than AC250V of the charged object measurement without damage to the instrument,;Thermocouple probe can contact the heating body has the limit, the order should be put forward to measure the temperature range, to provide the corresponding protective material

JK-XC multi-channel temperature tester is suitable for the lighting, household appliances, electric motors, electric appliances, temperature control devices, transformers, thermal protectors, and other manufacturers of the industry and the quality inspection departments of multi-point temperature field detection.

Notes:

① The T J K in the table is the type of the probe and the temperature range is:

J Graduation-100~750±℃(0.3%rdg+1.0℃) Resolution0.1℃

K Graduation100~1200±℃(0.3%rdg+1.0℃)   Resolution0.1℃

T Graduation-100~400±℃(0.3%rdg+1.0℃)     Resolution0.1℃

②curve:Refers to the instrument does not software in the LCD screen real-time display temperature curve and stored in the instrument, The curve can be permanently stored in the computer through the communication interface

③deta :The software will not instrument temperature value in number, number, acquisition time and stored in the corresponding temperature instrument,

The data can be permanently stored in the computer through the communication interface. The factory is equipped with a 2 meter long road temperature (temperature resistance of 250 degrees Celsius) wire, if you need to be extended by the calculation of 10 yuan / m! With the K line to prevent sudden power failure impact test, it is recommended with UPS power supply


Model

JK-8U

JK-8A

JK-8C

Input type

Sensor: nickel chromium nickel silicon (K) thermocouple (type T, type J can be specially made)

Test range

1、test temperature range:-100℃~1000℃;
2、test accuracy: 0~1000℃:±(reading ×0.5%+1)℃,-100~0℃:±(reading×0.5%+2)℃;

Channel number 8

8 road 16 road 24 road 32 road 40 road 48 road 64 road

display

LCD 240*128 Dot matrix screen

Single screen display

32 road

U disk storage

yes

no

yes

Live test

300V

Internal storage

yes

no

yes

Alarm mode

Digital flicker

Anti-interference

Anti high frequency interference

communication interface

USB

RS232

software kit

2014editionlatest

2000edition

Permissible environmental conditions

1、Power supply:AC 220V±10%,50Hz±2%;
2、Working environment:working temperature:-20-70℃,relative humidity:20%-90%;

Outside size

:(length×width×height)36cm×26cm×16cm machine weightabout 5kg